Gas Analyzer / On-Line Process
>> PRODUCT MAKER :    
 
Model : 3000HM
 
¡Ü HFID Total Hydrocarbon Analyzer
¡Ü ºü¸¥ ÀÀ´ä ¼Óµµ (T95) : 1.5s
¡Ü Stainless steel ÀçÁúÀÇ sampling system
¡Ü Catalytic air purifier ¿Í air compressor°¡ ±âº» »ç¾çÀ¸·Î ³»Àå
¡Ü Auto-Start up, Auto-Ignition, Auto-Cal, Auto-Range,
Health checks, Alarms
¡Ü 19" rack mounting 3U high.
     
 
 
Detector
Flame ionisation detector with cylindrical collector, flame ignitor and sensor.

Ranges
0-4, 10, 40, 100, 400, 1000, 4000, 10000 ppm
Propane equivalence, plus auto-range.

Response time
1.5s (5% to 95%).

Bypass flow sensitivity
Less than 2% from 1 to 3 l/min.

Accuracy and repeatability
Better than ¡¾1% range or ¡¾0.2 ppm whichever is greater.

Zero Drift
Less than 0.1 ppm in 8 hours.

Noise
Less than ¡¾0.1 ppm or ¡¾0.1% of range.

Linearity
¡¾2% of point or ¡¾0.5% FSD on all ranges down to 10% of range.

Oxygen synergism
100 ppm C3H8 in air, changed to 100 ppm C3H8 in N2 has less than 2% effect when using H2He fuel.

Display
240¡¿64 pixel LCD display with back-light shows concentration, units and gas in large clear characters.

Remote control
All functions by 'AK' protocol using RS232 port, Range, gas path, detector mode, calibration and standby by contact closure or TTL/CMOS logic levels.

Outputs
0-10 Vdc and isolated 4-20 mA analogue outputs of
displayed range. Range analogue output of 1 to 8 Vdc.

Power
Dual voltage 115 Vac or 230 Vac ¡¾15% switchable on the rear panel. 600 VA maximum during warm-up

Options
H2 Fuel for ambient monitoring applications.

Environment
5-40¡É ambient temperature. 95% R.H. non-condensing.

Dimensions
19" rack mounting 3U high.(133.5mm)
Depth behind mounting face 525mm.

Services
H2 fuel at 60 ml/min, or H2He fuel at 180 ml/min.
Propane (C3H8) or Methane(CH4) in air for span calibration.
Zero grade air (or the internal zero air) for zero calibration.