Model
: 3000HM |
|
|
¡Ü HFID Total Hydrocarbon Analyzer ¡Ü ºü¸¥ ÀÀ´ä ¼Óµµ (T95) : 1.5s ¡Ü Stainless steel ÀçÁúÀÇ sampling system ¡Ü Catalytic air purifier ¿Í air compressor°¡ ±âº» »ç¾çÀ¸·Î ³»Àå ¡Ü Auto-Start up, Auto-Ignition, Auto-Cal, Auto-Range, Health checks, Alarms ¡Ü 19" rack mounting 3U high. |
|
|
|
|
|
|
|
Detector Flame ionisation detector with cylindrical collector, flame ignitor and sensor.
Ranges 0-4, 10, 40, 100, 400, 1000, 4000, 10000 ppm Propane equivalence, plus auto-range.
Response time 1.5s (5% to 95%).
Bypass flow sensitivity Less than 2% from 1 to 3 l/min.
Accuracy and repeatability Better than ¡¾1% range or ¡¾0.2 ppm whichever is greater.
Zero Drift Less than 0.1 ppm in 8 hours.
Noise Less than ¡¾0.1 ppm or ¡¾0.1% of range.
Linearity ¡¾2% of point or ¡¾0.5% FSD on all ranges down to 10% of range.
Oxygen synergism 100 ppm C3H8 in air, changed to 100 ppm C3H8 in N2 has less than 2% effect when using H2He fuel.
Display 240¡¿64 pixel LCD display with back-light shows concentration, units and gas in large clear characters.
Remote control All functions by 'AK' protocol using RS232 port, Range, gas path, detector mode, calibration and standby by contact closure or TTL/CMOS logic levels.
Outputs 0-10 Vdc and isolated 4-20 mA analogue outputs of displayed range. Range analogue output of 1 to 8 Vdc.
Power Dual voltage 115 Vac or 230 Vac ¡¾15% switchable on the rear panel. 600 VA maximum during warm-up
Options H2 Fuel for ambient monitoring applications.
Environment 5-40¡É ambient temperature. 95% R.H. non-condensing.
Dimensions 19" rack mounting 3U high.(133.5mm) Depth behind mounting face 525mm.
Services H2 fuel at 60 ml/min, or H2He fuel at 180 ml/min. Propane (C3H8) or Methane(CH4) in air for span calibration. Zero grade air (or the internal zero air) for zero calibration. |
|
|