Gas Analyzer / On-Line Process
>> PRODUCT MAKER :    
 
Model : 4000VM
 
¡Ü Heated Vacuum chemiluminescent NOX Analyzer
¡Ü Raw exhaust gas ÃøÁ¤
¡Ü Heated reaction chamber ³»Àå
¡Ü Automatic microprocessor operation
¡Ü Dry vacuum pump ³»Àå
¡Ü No, NO2, NOX ÃøÁ¤ ¹× NH3 ÃøÁ¤ °¡´É (option)
     
 
 
Detector
Chemiluminescent.
NO + O3 ¡æ NO2 + O2 + photon
Cooled Photomultiplier Sensor

Ranges
0-4, 10, 40, 100, 400, 1000, 4000, 10000 ppm plus auto-range

Response time
1.5s (5% to 95%)

Bypass flow sensitivity
Less than 1% from 1 to 3 l/min.

Accuracy and repeatability
Better than ¡¾1% range or ¡¾0.2 ppm which ever is the greater.

Temperature effect
Zero: Less than ¡¾0.5 ppm from 5¡É to 35¡É
Span: Less than ¡¾ 0.33%/¡É

Noise
Less than ¡¾0.5% ppm or ¡¾0.1% of range.

Linearity
¡¾2% of point or ¡¾1% FSD on all ranges down to 10% of range.

NO2 to no converter
95% efficient requiring periodic checking (use Signal NOXGEN tester). Converter Temperature adjustable from the front panel.

Ozoniser
Generates O3 from air with negligible NO production
Air must have less than -12¡É dew point

Converter efficiency
Greater than 95%

Display
240¡¿64 pixel LCD display with back-light shows concentration, units and gas in large clear characters.

Remote control
All functions by 'AK' protocol using RS232 port.

Outputs
0-10 Vdc and isolated 4-20 mA analogue outputs of display range.

Power
Dual voltage 115 Vac or 230 Vac ¡¾15% switchable on the rear panel.

Options
High corrosion resistant dry vacuum pump. Specify mains
voltage required. Pump Control Adaptor. NO2 Measurement.

Environment
5-40¡É ambient temperature.
95% R.H. non-condensing.

Dimensions
19" rack mounting 3U high. (133.5 mm) Depth behind mounting face 590 mm. Protrusion in front of mounting face 45 mm.

Weight approx. 35kg

Services
Air at 520 ml/min with a dew point of less than -12¡É
for the ozoniser. NO in N2 diluent for span calibration.
100% N2 for zero calibration.0 psi maximum.