Gas Analyzer / Portable Gas Analyzers
>> PRODUCT MAKER :    
 
Model : G4010
 
¡Ü Galvanic cell ¹æ½Ä ÈÞ´ë¿ë »ê¼ÒºÐ¼®±â
¡Ü °¡º±°í ưưÇÑ Àü¿ë ÄÉÀ̽º¶ó ÈÞ´ëÇϱ⠿ëÀÌ
     
 
 
Display : Multi digit LCD - character height 12.7mm

Display ranges
1) E type cell-
Display range 0.00% to 100.0%.
Resolution: 0.1% from 10 to 100%; 0.01% below 10%.
2) N type cell -
Display range 0ppm to 50%.
Resolution: 0.1% from 10% to 50%; 0.01% from 0.50% to 9.99%;10ppm from 500ppm to 4999ppm; 1ppm from 0ppm to 499ppm Minimum usable reading 100ppm

Analogue outputs
0 to 1 volt. Full scale point continuously variable over the following range:
E type cell version: 5% to 100%
N type cell version: 50ppm to 50%

Alarms (Visual display only)
Two ?each programmable for concentration, function (high, low and off), hysteresis

Stability : Better than 2% of full-scale per month

Cell life : E type - up to 5 years, N type - up to 18 months

Sample connections
Inlet and outlet: captive seal compression fittings suitable for 0.25 inch (or 6mm) o/d tube
Sample flow Between 100 and 300 ml/min for optimum operation. Max. 1 l/min

Sampling pressure
Pump off: Maximum Inlet pressure 3 bar gauge
Pump on: Minimum inlet pressure Atmospheric

Sample vent pressure
Atmospheric (for quoted accuracy) Cell must not be subjected to rapid pressure changes

Sample temperature : 5 to +40 ¡É (non condensing)

Speed of response : T90 - variable depending on sensor and concentration. Approximately 3s at % levels and 20s at ppm levels. The ppm figure assumes that the sensor is purged down to the ppm concentration of interest.

Ambient temperature : 0 to 40¡É - continuous, -5 to + 55 ¡É - intermittent

Power supply : Nicad rechargeable cell pack. Operating time at least 10 hours with pump off and 6 hours with pump on.
The instruments can also be AC powered from 100v to 240v 50/60Hz. When the AC power is applied the battery is also charged.