|
Model
: S4 AURORA |
|
|
Paramagnetic Oxygen Analyser
combustion testing, inert atmospheres, medical and safety for humans, internal combustion engine exhaust, ETC
- Touch-screen colour display - 0.1% O2 resolution - 0.1% linearity |
|
|
|
|
|
|
|
DETECTOR TYPE: Paramagnetic. ¡®Dumb bell¡¯ servo driven. RANGES: - Range A: 0-5%, 0-10%, 0-25%. or - Range B: 0-100% or - Range C: 95-100% NOISE: 0.005% O2. LINEARITY: +/- 0.1% O2 REPEATABILITY: < +/- 0.01% O2 ZERO DRIFT: <+/- 0.2% O2 per 30 days RESPONSE TIME: < 2.5s FLOW EFFECT: 1% change in reading with flow change of 0.5 l/min to 3 l/min into atmospheric pressure. TEMPERATURE EFFECT: - 0.005% O2 change on Zero reading with temperature change of 5 to 40 OC. - 0.015% O2 change on span reading with temperature change of 5 to 40 OC. TILT EFFECT: 0.01% O2 per degree of tilt. EXTERNAL MAGNETIC EFFECT: A mass of soft magnetic material anywhere on the enclosure will cause less than 0.1 % O2 change in reading. CONSTRUCTION: Wetted parts, 316 Stainless steel and Quartz DIMENSIONS: 19¡± X 133.3mm X 550mm Weight: 12 Kg POWER REQUIREMENTS: 110VAC - 250VAC or 24VDC. 70 Watts. OUTPUTS: 0-10 volts, RS232, Ethernet (remote software is included), Optional 4-20 mA. |
|
|